Model PD690
4 1/2 Digit Universal Process Meter
Model PD690 is a high performance, easy to use, industrial-grade
microprocessor digital process meter. It accepts all the standard process
signals: 4-20 mA, 1-5 V, 0-5 V and 0-10 V and displays these signals in
engineering units on a 4 1/2 digit display. The display includes an extra zero
which may be used to handle numbers up to 199,990.
- Easy Single Button Scaling with Lockout Feature
- 4-20 mA, 0-5 V, or 0-10 V Field Selectable Inputs
- 24 V Transmitter Power Supply Standard
- Stand Alone Scaling
- Steady, Accurate Display to +/- 199,990
- 11 point Linearization
- Square Root Extraction
- 4 1/2 Digit + Extra Zero
- NEMA 4X Front Panel
- 2 or 4 Relay = 4-20 mA Output Options
Stand alone scaling allows the PD690 to be scaled without applying an input.
Simply select internal calibration (I-CAL) mode and set desired display for the
4mA and 20 mA value. To scale inputs such as 0 -5 V, 1-5V or 0-10V without
applying a signal, first complete an initial calibration.
4 Visual Alarms Standard
- Front panel alarm status indication
- Perfect for 'Visual indication only' applications
- any combination of high or low set points
- 0-100% dead-band adjustment
Big bright steady display for big numbers
- 4 1/2 digit + extra zero display
- Displays up to +/- 199,990
- 0.56" high efficiency red LED's
- Steady and quick responding display
Powers the loop
The 24 Volt transmitter power supply is standard on the PD690. This feature
saves time and money by simplifying wiring and eliminating the cost of an
external transmitter power supply.
NEMA 4X Front Panel
Wet, dirty and dusty environments don't bother the PD690's NEMA 4X front
panel so it can be installed in almost any panel in the plant
Linearization
- The linearization feature has two modes of operation:
- The 11 point user calibration displays non-linear signals
- The square root extraction displays the flow rate from a differential
transmitter.
- 11 point user calibration
- input up to 11 calibration points
- perfect for non-linear signals
Square Root Extraction
- Displays flow rates
- user selectable low flow cutoff
- Only two calibration points required
- accurate to +/- 0.1% FS +/- 1 count from 10-100% of flow
- The square root extraction feature displays flow rate by extracting
the square root of a signal from a differential pressure transmitter.
The user selectable low flow cut off feature gives a reading of zero
when the flow rate drops below a user-set point.
NEW 18-36 Vdc POWERED
The PD690 Universal Process Meter can now be powered from 18-86 Vdc. This new
option provides up to 500 Vdc of isolation. All the features and options
available on the AC powered meters are available on the DC powered meters.
Note: 18-86 VDC powered PD690's are not UL listed.
Remember
Every PD690 comes with 4 visual alarms standard! The PD690 universal process
meter can be equipped with 2 or 4 relay contacts and a isolated 4-20 mA
transmitter output. These options may be combined in any configuration to
satisfy a wide variety of applications. A fully loaded model PD690-3 includes 4
relays and a isolated 4-20 mA transmitter output.
Two or four Relays
- 2 or 4 SPDT, 2A @240Vac resistive
- Automatic or automatic + manual reset
- Fail-safe action and auto initialization
- 0-100% dead-band adjustment
Isolated 4-20 mA output
- 4-20mA or 20-4 mA output
- 500 Vdc isolation
- With or without relay contacts
Specifications-Basic Meter
- Inputs
: Field Selectable,4-20 mA, 1-5 V, 0-5 V, 0-10 V.
- Display
: Bright, Large, 0.56" (14.2mm) high efficiency red LED. 4
1/2 digits + extra zero; +/- 1999(0),(0) may be switched on to display
199,990.
- Front Panel
: NEMA 4X, panel gasket provided.
- Calibration
: Maybe calibrated using internal calibration (I-CAL) or
by applying an external calibration signal (E-CAL).
- Calibration Range
: 4 mA (1V) input may be set to display anywhere in
range of meter. 20 mA (5V) may be set anywhere above or below 4MA (5V) may be
set anywhere above or below 4 mA input.
- Lockout
: Jumper 3 at rear of instrument restricts modification of
calibration values.
- Loop Power
: Isolated 24 VDC at 20mA regulated. Noise less than 10 mV
p-p. Max loop resistance of 1200 ohms. Available for either signal input or
4020 mA output option, but not both.
- Hold Reading
: Connect terminals HLD and COM.
- Accuracy
: 0.05% of calibrated span.
- Input Impedance
: Voltage ranges, greater than 100 Kohms: Current
range,100 ohms.
- Power
: 115 VAC or 230 VAC +/- 10%, 50/60 Hz, 10 VA.
- Environmental
: Operating temperature range: 0 to 65 deg C. Storage
Temperature range : -40 to 85 deg C. Relative Humidity : 0 to 90%
non-condensing.
- Enclosure
: 1/8 DIN, high impact plastic, UL 94V-0.
- Connections
: Removable screw material block (provided).
- Alarm Points
: 4,any combination of high or low alarms.
- Alarm Status Indication
; Front Panel LED
- Alarm Dead-band
: 0-100%, User Selectable.
- UL File Number
: E160849; 508 Industrial Control Equipment.


Relays
- Rating
: 2 or 4 relays; 2 Amp SPDT (form C). The contacts are rated 2
Amp @ 30 VDC or 2 Amp @ 250 VAC resistive load; 1/14 HP @ 125/250 VAC for NC
contacts and 1/10 HP @ 125/250 VAC for NO contacts, inductive load.
- Reset
: User Select.
- 1. Automatically when the input passes the reset point.
- 2. Automatically + manually (via user supplied switch or front panel ACK
button).
- Fail-safe Operation
: The relay coils are energized in the non-alarm
condition. In the case of a power failure , the relays will go to the alarm
state. May be disabled with internal jumper.
- Auto Initialization
: When power is applied to the meter, the relays
will always reflect the state of the input to the meter.
- Dead band
: 0-100 %, User Selectable.
Isolated 4-20 mA Transmitter Output:
- Calibration Range
: Anywhere in range of meter, 501 count minimum span.
- Output Loop Resistance
: 1500 ohms max. using a 35V external power
supply. 500 ohms max. using the built-in loop power supply. Built-in loop
power supply available for either 4-20 mA output or signal input, but not
both.
- Accuracy
: +/- 0.1% FS, +/- .004 mA
- Isolation
: 500 VDC or peak AC, input-to-output or
input/output-to-power line.
- External Loop Power Supply
: 35 V max.
Hot tip - Green Lee (800 435 0786) has a 1/8th DIN panel punch
part # 60068 for the NEMA 4X bezels.
| Order number |
Description | Repacement |
| PD690-3-N |
No options | |
| PD690-3-14 |
2 Relays | |
| PD690-3-15 |
4-20mA Output | |
| PD690-3-16 |
2 Relays + 4-20 mA output | |
| PD690-3-17 |
4 Relays | |
| PD690-3-18 |
4 Relays + 4-20 mA output | |
| A -4 may be substituted for the -3 to call out meters
powered from 230Vac at no extra charge. | |
Prices subject to change.